Agilent Cary Eclipse fluorescence spectrophotometer
URI | http://vocab.nerc.ac.uk/collection/L22/current/TOOL1222/ |
---|---|
Within Vocab | SeaVoX Device Catalogue |
Preferred Label | Agilent Cary Eclipse fluorescence spectrophotometer |
Definition | A laboratory instrument that measures the amount of light at discrete pre-determined wavelengths that is transmitted or emitted by a sample. It can be operated with measurement modes for fluorescence, phosphorescence, chemiluminescence and bioluminescence. The instrument uses Czerny-Turner 0.125 m monochromators with a 190-1100 nm wavelength range and a fixed selectable spectral bandwidth from 1.5 to 20 nm. It uses a full spectrum Xe pulse lamp single source and dual R928 photomultiplier tubes. The instrument has scan rates from 0.01 to 24000 nm/min, and a maximum measurement rate of 80 data points per second. It also has room light immunity when in fluorescence mode. |
Broader |
L05:
fluorometers |
Related |
B75:
Agilent Technologies UK Limited - Life Sciences & Chemical Analysis Group |
Note | accepted |
Deprecated | false |
Alternative Label | Agilent Cary Eclipse |
Version | 1 |
Has Current Version | http://vocab.nerc.ac.uk/collection/L22/current/TOOL1222/1/ |
Authored On | 2018-04-10 17:26:52 |