Concept
Idronaut Ocean Seven 316 plus CTD
URI | http://vocab.nerc.ac.uk/collection/L22/current/TOOL1306/ |
---|---|
Within Vocab | SeaVoX Device Catalogue |
Preferred Label | Idronaut Ocean Seven 316 plus CTD |
Definition | An underwater CTD probe designed for on-line profiling or used in self-recording applications such as on moorings or AUVs. It comprises a piezoresistive pressure transducer, a platinum resistance thermometer, a platinum seven-ring quartz conductivity cell, a pressure-compensated polarographic oxygen sensor, a pH glass sensor and solid gel pH reference probe. The conductivity cell employs a large diameter (8 mm) and a short length (46 mm) which facilitates self-flushing and prevents clogging. The conductivity cell design does not require platinum black deposition and can be cleaned in the field without re-calibrating. The probe can accommodate up to 16 auxiliary sensors. The probe can be interfaced with a telemetry deck unit for real-time data acquisitions. It is depth rated to 7000 dbar. Pressure accuracy: 0.01 - 0.05 % FS. Temperature accuracy: 0.002 deg C. Conductivity accuracy: 0.003 mS/cm. |
Broader |
L05:
CTD L05: water temperature sensor L05: salinity sensor L05: dissolved gas sensors L05: pH sensors |
Related |
B75:
Idronaut |
Note | accepted |
Deprecated | false |
Alternative Label | Idronaut OS 316 plus |
Version | 1 |
Has Current Version | http://vocab.nerc.ac.uk/collection/L22/current/TOOL1306/1/ |
Authored On | 2019-02-06 12:10:27 |
Alternate Profiles