Concept
PerkinElmer Tri-Carb 2100TR low activity liquid scintillation analyser
URI | http://vocab.nerc.ac.uk/collection/L22/current/TOOL1013/ |
---|---|
Within Vocab | SeaVoX Device Catalogue |
Preferred Label | PerkinElmer Tri-Carb 2100TR low activity liquid scintillation analyser |
Definition | A benchtop liquid scintillation analyser for detecting small amounts of alpha, beta and gamma radioactivity. It features a Multichannel Analyzer with an effective resolution of 1/10 keV and an extended dynamic quench range. Sample capacity is either 408 standard 20 mL vials, or 720 small 4 or 7 mL vials. The instrument includes a barium-133 low energy external standard source centered under the sample vial which eliminates the effects of volume variations. It has an energy range of 0-2000 keV and an operating ambient temperature range of 15-35 degC. This instrument was superseded by the 3100TR and subsequently the 3110TR. |
Broader |
L05:
liquid scintillation counters |
Related |
B75:
PerkinElmer |
Note | accepted |
Deprecated | false |
Alternative Label | Tri-Carb 2100TR liquid scintillation |
Version | 1 |
Has Current Version | http://vocab.nerc.ac.uk/collection/L22/current/TOOL1013/1/ |
Authored On | 2016-08-23 15:42:03 |
Alternate Profiles
Different views and formats: