Manchester Centre for Atmospheric Science Hygroscopic Tandem Differential Mobility Analyser
URI | http://vocab.nerc.ac.uk/collection/L22/current/TOOL0352/ |
---|---|
Within Vocab | SeaVoX Device Catalogue |
Preferred Label | Manchester Centre for Atmospheric Science Hygroscopic Tandem Differential Mobility Analyser |
Definition | The Hygroscopic Tandem Differential Mobility Analyser (HTDMA) measures the changes in aerosol size distribution in response to a change in relative humidity. It uses two Differential Mobility Analysers (DMAs), operated in series. The incoming air sample is dried to <10% relative humidity, and the first DMA selects a monodisperse distribution of a chosen size. The monodisperse subsample is humidified, and the second DMA, coupled to a particle counter, measures the size distribution of the humidified sample. |
Broader |
L05:
Aerosol physical characterisers |
Related |
B75:
University of Manchester, School of Earth, Atmospheric and Environmental Sciences |
Note | accepted |
Deprecated | false |
Alternative Label | Manchester CAS HTDMA |
Version | 1 |
Has Current Version | http://vocab.nerc.ac.uk/collection/L22/current/TOOL0352/1/ |
Authored On | 2010-10-01 10:54:41 |
Alternate Profiles