Concept
Thermo Fisher Scientific iCAP TQ inductively coupled plasma mass spectrometer
URI | http://vocab.nerc.ac.uk/collection/L22/current/TOOL2138/1/ | ||
---|---|---|---|
Within Vocab | SeaVoX Device Catalogue | ||
Alternative Labels | Thermo Scientific iCAP TQ ICP-MS | ||
Definition | A benchtop triple quadrupole (TQ) inductively coupled plasma mass spectrometer (ICP-MS) with a four channel peristaltic pump, three plasma gas flow controllers, and four QCell mass flow controllers. The iCAP TQ utilises triple quadrupole technology which allows the analyte signal to be isolated from interferences, enabling the analysis of complex or high-matrix samples. The high frequency (4 MHz) quadrupole mass analyser has pre and post filters for isolation of wanted ions. The instrument features Peltier cooled high purity quartz or perfluoroalkoxy alkane (PFA), and low volume, baffled cyclonic or double pass spray chambers to efficiently filter out larger aerosol droplets for improved plasma stability. A reaction finder method development assistant aids easy method development. The plasma system is designed to rapidly adapt to changing matrices and provide robustness for challenging samples such as highly volatile organic solvents. The argon ICP ion source has a digital, solid state radiofrequency generator, and dynamic frequency impedance matching the plasma at 27 MHz. The iCAP TQ has applications in trace element analysis of solid or liquid (particularly sediment or sea water) samples. It has a nebuliser default flow rate of 400 uL/min. | ||
Date | 2025-02-14T11:24:53 | ||
Identifier | SDN:L22::TOOL2138 | ||
Note | accepted | ||
Is Version Of | TOOL2138 | ||
version | 1 | ||
Broader | L05:LAB15 |
inductively-coupled plasma mass spectrometers | |
Related | B75:ORG00458 |
Thermo Fisher Scientific | |
L35:MAN0035 |
Thermo Fisher Scientific (Thermo Scientific) |
Alternate Formats
Other formats for this page:
RDF/XML Turtle JSON-LDAlternate Profiles
Other views of this page:
Alternate Profiles ?Different Media Types (HTML, text, RDF, JSON etc.) and different information model views, profiles, are available for this resource.
NVS html view ? Default NVS html view.