Concept
Marianda VINDTA 3C total inorganic carbon and titration alkalinity analyser
| URI | http://vocab.nerc.ac.uk/collection/L22/current/TOOL0481/1/ | ||
|---|---|---|---|
| Within Vocab | SeaVoX Device Catalogue | ||
| Alternative Labels | VINDTA 3C TIC and TA analyser | ||
| Definition | The Versatile INstrument for the Determination of Total inorganic carbon and titration Alkalinity (VINDTA) 3C is a laboratory alkalinity titration system combined with an extraction unit for coulometric titration, which simultaneously determines the alkalinity and dissolved inorganic carbon content of a sample. The sample transport is performed with peristaltic pumps and acid is added to the sample using a membrane pump. No pressurizing system is required and only one gas supply (nitrogen or dry and CO2-free air) is necessary. The system uses a Metrohm Titrino 719S, an ORION-Ross pH electrode and a Metrohm reference electrode. The burette, the pipette and the analysis cell have a water jacket around them. Precision is typically ± 1 umol/kg for TA and/or DIC in open ocean water. | ||
| Date | 2012-03-19T13:44:14 | ||
| Identifier | SDN:L22::TOOL0481 | ||
| Note | accepted | ||
| Is Version Of | TOOL0481 | ||
| version | 1 | ||
| Broader | L05:LAB12 |
titrators | |
| Related | B75:ORG00584 |
MARIANDA | |
Alternate Formats
Other formats for this page:
RDF/XML Turtle JSON-LDAlternate Profiles
Other views of this page:
Alternate Profiles ?Different Media Types (HTML, text, RDF, JSON etc.) and different information model views, profiles, are available for this resource.
NVS html view ? Default NVS html view.