Concept
WETLabs ECO-FLNTU(RT)D combined fluorometer and turbidity sensor
URI | http://vocab.nerc.ac.uk/collection/L22/current/TOOL1531/2/ | |
---|---|---|
Within Vocab | SeaVoX Device Catalogue | |
Alternative Labels | WETLabs ECO-FLNTU(RT)D | |
Definition | This optical sensor is available in combinations of backscattering, turbidity, and fluorescence measurements. It records in real-time and does not store data. ECOs feature optional active anti-fouling and internal batteries for long-term deployments. This instrument has a user-selectable sample rate up to 8 Hz The fluorometer can typically measure pigment concentrations in the range 0-75 ug/l, with a sensitivity of 0.037 ug/l, at wavelengths of 470 or 695 nm. The turbidity sensor can measure within the range 0-200 NTU, with a sensitivity of 0.098 NTU, at a wavelength of 700 nm. The instrument is stable over a temperature range of 0-30 degC and is rated to a depth of 6000 m. | |
Date | None | |
Identifier | TOOL1531 | |
Note | accepted | |
Is Version Of | TOOL1531 | |
version | 2 |