A temperature and humidity probe. It is used for long-term, unattended applications. The probe uses a capacitive sensor (HYGROMER IN-1) to measure relative humidity and platinum resistance thermometer (PT100) to measure temperature. The probe includes a polyethylene or Teflon filter (40um) to protect the sensor from dust and particles. The (L) notation indicates that this sensor is supplied with a user-specified cable length. Temperature accuracy is 23 degC (+/- 0.1 degC). Humidity accuracy is +/- 0.8%.