SDN:L22::TOOL1236
Shimadzu GC-2010 series
Shimadzu GC-2010 series
Shimadzu GC-2010 (capillary) gas chromatograph series
accepted
Bench-top gas chromatographs. They separate and analyse gas mixtures using a capillary column. Features a temperature controlled oven, up to 3 temperature controlled injection units, up to 4 detectors and a flow control unit. Detectors used are Flame Ionization (FID), Thermal Conductivity (TCD), Electron Capture (ECD), Flame Photometric (FPD) and Flame Thermionic (FTD). The standard GC-2010 was developed in 2000 while the GC-2010 Plus was developed in 2009 and features improved noise reduction and detector capability to improve sensitivity. For example, the minimum detected quantity for flame photometric detection is 0.2 pg P/s, 4 pg S/s (std) and 55 fg P/s, 3 fg S/s (Plus). Maximum operating pressure: 970 kPa. Maximum flow rate: 1200mL/min.