Concept
PerkinElmer Tri-Carb 3100TR low activity liquid scintillation analyser
URI | http://vocab.nerc.ac.uk/collection/L22/current/TOOL0820/2/ | |
---|---|---|
Within Vocab | SeaVoX Device Catalogue | |
Alternative Labels | Tri-Carb 3100TR liquid scintillation | |
Definition | A benchtop liquid scintillation analyser for detecting small amounts of alpha, beta and gamma radioactivity. It features a Multichannel Analyzer with an effective resolution of 1/10 keV and an extended dynamic quench range. Sample capacity is either 408 standard 20 mL vials, or 720 small 4 or 7 mL vials. The instrument does not allow for the interchange of large and small sample vials. The instrument includes a barium-133 low energy external standard source centred under the sample vial which eliminates the effects of volume variations. It has an energy range of 0-2000 keV and an operating ambient temperature range of 15-35 degC. This instrument was superseded by the 3110TR. | |
Date | None | |
Identifier | TOOL0820 | |
Note | accepted | |
Is Version Of | TOOL0820 | |
version | 2 |
Alternate Formats
Other formats for this page:
RDF/XML Turtle JSON-LDAlternate Profiles
Other views of this page:
Alternate Profiles ?Different Media Types (HTML, text, RDF, JSON etc.) and different information model views, profiles, are available for this resource.
NVS html view ? Default NVS html view.